requires a Station Identification (ID code) so that it can clearly be

 

 

 

 

identified.

 

 

 

 

1.

Select the menu item REMOTE ID in Menu 2.

 

 

 

 

2.

Enter an ID with the UP or DOWN Scroll Keys

 

 

 

 

 

(possible values 0 – 999).

 

 

 

 

3.

Proceed to the next menu item with the Radio Button ┌ 1 ┐ (back)

 

 

 

 

 

or ┌ 2 ┐ (forward) or exit the menu with Radio button ┌ 3 ┐ (EXIT).

 

 

Note

Depress Radio Button ┌ 3 ┐ to exit the menu item without saving any

 

 

 

 

changes (EXIT).

 

 

 

 

Factory Control Check ( FCC )

 

 

 

 

This function checks temperature accuracy of the Rework Station

 

 

 

 

and allows modifications if necessary. To perform the “FCC” function,

 

 

 

 

the soldering tip temperature must be measured using an external

 

 

 

 

temperature measuring instrument, (WA2000) and appropriate type

 

 

 

 

“K” temperature measuring tip. ( See Accessories List on page 18 )

 

 

 

 

Control Check at 212 °F / 100 °C

 

 

 

 

1.

Connect the type “K” thermocouple, into the external temperature

 

 

 

 

 

measuring instrument, (WA2000).

 

 

 

 

2.

Select the menu item FCC in Menu 2.

 

 

 

 

3.

Depress the DOWN Scroll Key. Control point 212 °F / 100 °C is

 

 

 

 

 

 

 

 

 

selected.

 

 

 

 

 

The soldering tip is now heated to 212 °F / 100 °C.

 

 

 

 

 

The Heater Control Indicator flashes ( ) (21) as soon as the

 

 

 

 

 

temperature is constant.

 

 

 

 

4.

Compare the temperatures indicated by the meter with the

 

 

 

 

 

indications in the display.

 

 

 

 

5.

Use the UP or DOWN Scroll Key to set the difference between the

 

 

 

 

 

value indicated on the external meter and the value indicated on

 

 

 

 

 

the Rework Station, (shown in small script in lower part of the

 

 

 

 

 

display). Maximum temperature adjustment: ± 72 °F (± 40 °C).

 

 

 

 

 

Example:

 

 

 

 

 

Display 212 °F, external meter 210 °F: setting S 2

 

 

Note

 

Display 212 °F, external meter 214 °F: setting T 2

 

 

Depress Radio Button ┌ 3 ┐ to exit the menu item without saving

 

 

 

 

changes (EXIT).

 

 

 

 

6.

Depress Radio Button ┌ 2 ┐ (Set) to confirm the value.

 

 

 

 

 

The temperature deviation is now reset to 0. Factory Control

 

 

 

 

 

Check at 212 °F / 100 °C is now complete.

 

 

 

 

7.

Exit Menu 2 by depressing Radio Button ┌ 3 ┐ twice.

Control Check at 842 °F / 450 °C

1. Connect the type “K” thermocouple, into the external temperature measuring instrument, (WA2000).

Page 15
Image 15
Cooper Lighting Radio manual Changes Exit, Factory Control Check FCC, Control Check at 212 F / 100 C