The following examples show some of the typical conditions occurring during the installation:

Figure 34 - Example Under Exposure: Too Dark

Under-exposure:

To correct this result it is recommended to change the following parameters in their order of appearance:

1.increase the Exposure Time (x 10 µs)

2.increase the Gain

 

In general, a longer exposure time corresponds to a lighter image

 

but is susceptible to blurring due to code movement. Exposure time

 

is also limited by the Internal Lighting mode parameter. Longer

 

esposure times can be set if the power strobe level is lowered.

NOTE

High gain settings may produce a grainy image that may affect the

decoding process.