BM122.qxp 9/10/07 2:24 PM Page 23

GB

TAKE NOTE

Caution: The sleeve size only covers a size range of 24 to 32 cm in diameter and is only to be used by adults.

Do not use the measuring unit if your arm is injured.

If the measuring appliance should continue inflating, open the sleeve immediately or press the START/STOP button.

By pressing START/STOP during inflation, the inflation process is stopped and the air is immediately released from the sleeve.

Should by any means pain occur during the measurement, stop immediately by pushing the START/STOP button or by opening the sleeve.

TECHNICAL DATA

Measuring procedure

Oscillometrical

Measurement range

Pressure 0 ~ 300 mmHg, Pulse 40 ~ 199 beats per minute

Measurement accuracy

Pressure +/- 3mmHg, Pulse max. +/- 5%

Inflation

Electric rolling pump

Deflation

Magnetic valve

Display

Systole, Diastole, pulse value, year/date/time, memory database

Memory Database

with 99 memory spaces

Automatic switch off

1 minute after last pressing of any button

Weight

Approx. 265g with batteries

Sleeve size

Arm diameter of about 24 to 32 cm

Operating temperature range

+10°C to +40°C, <85% relative humidity

Storage temperature range

-20°C to +70°C, <85% relative humidity

Power supply

6V DC 4 x 1,5 V R6/LR6/AA alkaline batteries

Battery life

300 measurements (2 per day)

Measures

80 x 120 x 40 mm

 

- 23 -

Page 23
Image 23
Elta BM122 instruction manual Take Note, Technical Data