Rosemount 8712 / 8700 Series
STEP 5: GROUNDING
Use Table 6 to determine which process grounding option to follow for proper installation. The sensor case should be earth grounded in accordance with national and local electrical codes. Failure to do so may impair the protection provided by the equipment.
Table 6. Process Grounding Installation
|
| Process Grounding Options |
| ||
Type of Pipe | Grounding |
| Grounding | Grounding | Lining |
Straps |
| Rings | Electrode | Protectors | |
Conductive | See Figure 10 |
| Not Required | Not Required | See Figure 11 |
Unlined Pipe |
|
|
|
|
|
Conductive Lined | Insufficient |
| See Figure 11 | See Figure 10 | See Figure 11 |
Pipe | Grounding |
|
|
|
|
Insufficient |
| See Figure 12 | See Figure 13 | See Figure 12 | |
Pipe | Grounding |
|
|
|
|
Figure 10. Grounding Straps or Grounding Electrode in Lined Pipe
Figure 11. Grounding with Grounding Rings or Lining Protectors
Grounding Rings or Lining
Protectors