5 Parts and Controls
5.1 Front Panel Buttons
See Figure3 .
The front panel buttonsare used to select units of measurement, access operat-
ing parameters, and alter operating parameters. The function of each button is
as follows:
C/Probe—Thisbutton selects u nits of degrees Celsius. In conjunction with the
Menu button,it selects the probe parameter menu.
F/Sample—This buttonselects units of degrees Fahrenheit. In conjunction
with the Menu button,it selects the sample parameter menu.
K/Comm—Thisbutton selects u nits of Kelvin.I n conjunction with the Menu
button,it selects the communication parameter menu.
Ω/Exit (Cal)—This button selects resistance in ohms. While editing a parame-
ter,it cancels the immediate operation and skips to the next parameter. If the
Exit buttonis pressed for more than one-half second the menu is exited. In
conjunction with the Menu button, it selects the calibration parameter menu.
Menu/Enter—Thisbutton allows one of the unit/menu buttons to select a
menu. When editing a parameter,it accepts the new value and skips to the next
operation.
Land R—When editing a numeric parameter, these buttons movebetween
digits.The selected digit flashes.
Uand D— When editing a parameter,th ese buttons increase or decrease the
valueof th e parameter or a selected digit.
15
5 Parts and Controls
Front Panel Buttons
C
PROBE
F
COMM
K
SAMPLE
W
EXIT
MENU
ENTER
84.9814 C1504THERMOMETER
READOUT
Figure3 1504 Front Panel