3 Specifications and EnvironmentalConditions

3.1 Specifications

Resistancerange 0to1M
Ω
,auto-ranging
Resistanceaccuracy, one
year
1
0to5k
Ω
:0.5
Ω
5k
Ω
to200 k
Ω
:0.01% (100 ppm) of reading
200k
Ω
to1M
Ω
:0.03% (300 ppm) of reading
Resistanceaccuracy, short
term
1
0to5k
Ω
:0.25
Ω
5k
Ω
to200 k
Ω
:0.005% (50 ppm) of reading
200k
Ω
to1M
Ω
:0.02% (200 ppm) of reading
Temperatureaccuracy,
typical
1,3, 4
0°C:0.002°C
25°C:0.002°C
50°C:0.004°C
75°C:0.010°C
100°C:0.020°C
(Using10K
Ω
thermistorsensor,
α
=0.04. Does not inclue probe uncer-
taintyor c haracterization errors.)
Temperaturecoefficient of
resistance
1
2.5ppm/°C
Resistanceresolution 0to10k
Ω
:0.01
Ω
10k
Ω
to100 k
Ω
:0.1
Ω
100k
Ω
to1M
Ω
:1
Ω
Temperatureresolution 0.0001°C(0.0001°F)
Probe Thermistor;also, 1k
Ω
RTD
Numberof input channels 1
Probeconnection Four-wirewith shield, 5-pin DIN connector
Maximumacceptable lead
resistance 100

Ω

Probecharacterizations Steinhart-Hartexponential polynomial
Callendar-VanDusen; R
0
,
α
,
δ
,and
β
Displayunits C(degrees Celsius)
F(degrees Fahrenheit)
K(Kelvin)
Ω
(resistancein ohms)
Display 8-digit,7-segment, yellow-green LED; 0.5 inch high characters
Clockaccuracy, typical 0.01%
Probeexcitation current 0to51k
Ω
:10
μ
A
51k
Ω
to1M
Ω
:2
μ
A
2Hz, alternating
Measurementperiod 1second
Digitalfilter Exponential,0 to 60 secinds time constant (user selectable)
9
3 Specifications and Environmental Conditions
Specifications