Communications RS-232serial standard,
IEEE-488(GPIB) optional, Conforms to IEEE-488.1, Capability: AH1,
SH1,T6, L4, DC1
OperatingTemperature
Range
Fullaccuracy: 13°C to 33°C
Absolute: 0°C to 55°C
ACPower 115VAC (±10%), 50/60 Hz, 10 W, nominal 1 A maximum
230VAC (±10%), 50/60 Hz, 10 W (optional)
Detachablepower cord
DCPower (optional) 10–14VDC (±10%), 1 A maximum
(220mA typical, normal mode; 120 mA typical, power saver mode)
Size 5.6in. (14.3 cm) wide x 7.1 in. (18.1 cm) deep x 2.4 in. (6.1 cm) high
Weight 2.2lbs. (1.0 kg.)
Safety OVERVOLTAGE(Installation) CATEGORY II, Pollution Degree 2 per IEC
1010-1
1
Accuracyspecifications apply within the recommended operating temperature range. Accuracy limits are
increasedby a factor of the temperature coefficient outside this range.
2
Short-termaccuracy includes nonlinearity and noise uncertainties. It does not include drift or calibration
uncertainties.
3
Thetemperature range may be limited by the sensor.
4
Temperatureaccuracy is for the 1504 only. It does not include probe uncertainty or probe characteriza-
tionerrors .A ccuracies are typical with 10 k
Ω
NTCthermistors.
3.2 Environmental ConditionsAlthough the instrumenthas been designed for optimum durability and trou-
ble-freeoperation, it must be handled with care. The instrument should not be
operated inan excessively dusty or dirty environment. Maintenance and clean-
ing recommendationscan be found in the Maintenance Section of this manual.
The instrument operates safely under thefollowing conditions:
•Operating temperature range: Absolute 0–55°C (32–131°F); [full accu-
racy 16–30°C (61–86°F)]
•Ambient relative humidity: maximum 80% for temperature < 31°C, de-
creasing linearly to 50% at 40°C
•Pressure: 75kPa–106kPa
•Mains voltage within ±10% of nominal
•Vibrations should be minimized
•Altitude less than 2,000 meters
•Indoor use only
1504 Thermometer Readout
User’s Guide
10