7.2.1 Setting the Address
The IEEE-488 busrequires that each device has a unique address. The default
address of the1504 is 22 but can be changed if necessary. The IEEE-488 ad-
dress of the 1504 is setwithin the Comm menu after the serial linefeed param-
eter.(This menu option will not appear if the IEEE-488 interface is not
installed). Press the Menu button (“SEt?” appears) then press the Comm but-
ton. The displaywill briefly indicate “SErIAL”, then “bAUd” and then display
the currentbaud rate.Press Enter severaltimes until “IEEE”appears.Thedis
-
play willbriefly indicate “AddreSS” and then display the current IEEE-488
address. Use the Uand Dbuttonsto change the number then press Enter.
7.2.2 Setting the Termination Character
The 1504 willnormally terminate transmissions from the IEEE-488 port with a
linefeed (newline)character. Some systems may require a terminating carriage
return instead. The termination character can bechanged if necessary. The ter-
mination character is set within the Comm menu after the IEEE-488 address
parameter.(This menu option will not appear if the IEEE-488 interface is not
installed).The display will briefly indicate “EOS” (end of string) and then dis-
play the currentsetting. Use the Uand Dbuttons to change th e termination
character then press Enter.
7.2.3 Time Stamp
Measurement data read from the GPIB interfacecan be stamped with the
time-of-day.For instructions on setting the time stamp and system clock see
Section7.1.3a bove.
7.3 Remote CommandsASCII commandsare used to instruct the 1504 to perform certain actions. Ta -
ble5 providesa complete list of commands. These commands can be used with
either the RS-232 or IEEE-488 interface.All commands sent to the 1504 must
be terminatedwith a carriage return or linefeed. Either upper or lower case let-
ters are accepted. Commands used to set a parameter are issued with the com-
mand header,an “=“ character, and the parameter value. For example,
U=C<EOS> sets the units to Celsius. (The symbol <EOS> represents the termi-
nation character.)Commands used to request data are issued with only the
command header.For example, T<EOS> causes the 1502A to return the most
recent measurement. Basic operations using commands are explainedin the fol-
lowingsub-sections.
7.3.1 Measurement Commands
The following commands relate to reading measurement data.
1504 Thermometer Readout
User’s Guide
28