Model 187 & 189
Users Manual
dB is calculated with the following formula:
dB = | | Vx | |
20 * log10 |
| | |
| |||
| | Vr |
•For dBm, Vr is the voltage across the reference resistance at 1 mW. For example, Vr would be
0.7746 V with a 600 Ω reference resistance.
•For dBV, the reference voltage (Vr) is 1 V.
Measuring DC Voltage
Set up the meter for dc voltage measurement as shown in Figure
Both AC and DC Voltage Measurements
When a dc volts function is selected, the meter can display ac and dc components of a signal separately or the combined ac + dc (rms) value.
To select separate ac and dc signal components:
•Press % once to display ac voltage in the primary display and dc voltage in the secondary display (ac over dc).
•Press % a second time to reverse the displays (dc over ac).
•Press % a third time to display the ac + dc rms value in the primary display. (FAST MN MX is unavailable in this state.)
•Press % a fourth time to return to the normal dc volts display.