Chapter 7

Specifications

Safety and Compliances

Maximum voltage between any terminal and earth ground.

1000 V dc or rms ac

Compliances - DUAL RATINGS

Certifications (Iisted and pending)

Surge Protection

WFuse Protection for mA or μA inputs

Fuse Protection for A input

Markings

Complies with IEC 1010-1 to 1000 V Overvoltage Category III, ollution Degree 2; and IEC 664-1 to 600 V Overvoltage Category

IV, Pollution Degree 2 *

CSA per standard CSA/CAN C22.2 No. 1010.1-92

UL per standard UL 3111 TÜV per standard EN 61010 Part 1-1993

8 kV peak per IEC 1010.1-92

0.44A (44/100 A, 440 mA), 1000 V FAST Fuse 11 A, 1000 V FAST Fuse

P, ),UL, TÜV and #

*OVERVOLTAGE (Installation) Categories refer to the level of Impulse Withstand Voltage protection provided at the specified Pollution Degree.

Overvoltage Category III equipment is equipment in fixed installations. Examples include switch gear and polyphase motors.

Overvoltage Category IV equipment is equipment for use at the origin of the installation. Examples include electricity meter and primary over-current protection equipment.

7-1