Electromagnetic Immunity

Specifications 9

Electromagnetic Immunity

Electromagnetic Immunity

The Fluke 190 series, including standard accessories, conforms with the EEC directive 89/336 for EMC immunity, as defined by EN-61326-1, with the addition of the following tables.

Scope Mode (10 ms/div): Trace disturbance with VPS200 voltage probe shorted

 

Table 1

 

No visible disturbance

 

E = 3V/m

 

Frequency range 10 kHz to 20 MHz

 

5 mV/div to 100 V/div

Frequency range 20 MHz to 100 MHz

 

200 mV/div to 100

V/div

Frequency range 100 MHz to 1 GHz

 

500 mV/div to 100

V/div *)

(*)With the 20 MHz Bandwidth Filter switched on: no visible disturbance.

With the 20 MHz Bandwidth Filter switched off: disturbance is max 2 div.

 

Table 2

Disturbance less than 10% of full scale

 

E = 3V/m

Frequency range 20 MHz to 100 MHz

 

10 mV/div to 100 mV/div

Test Tool ranges not specified in tables 1 and 2 may have a disturbance of more than 10% of full scale.

Meter Mode (Vdc, Vac, Vac+dc, Ohm and Continuity): Reading disturbance with test leads shorted

 

Table 3

Disturbance less than 1% of full scale

 

E = 3V/m

Frequency range 10 kHz to 1 GHz

 

500mV to 1000V , 500Ohm to 30 MOhm ranges

105

Page 113
Image 113