Fluke 196C Triggering on Edges, Open the Trigger on Edge menu, Select Free Run, jump to Noise

Models: 196C

1 118
Download 118 pages 62.3 Kb
Page 57
Image 57
Triggering on Edges

Triggering on Waveforms

5

Triggering on Edges

Triggering on Edges

If the signal is instable or has a very low frequency, use edge triggering to obtain full manual trigger control.

To trigger on rising edges of the input A waveform, do the following:

1

 

 

Display the TRIGGER key labels.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

When Free Run is selected, the test tool updates the screen even if there are no triggers. A trace always appears on the screen.

When On Trigger is selected, the test tool needs a trigger to display a waveform. Use this mode if you want to update the screen only when valid triggers occur.

When Single Shot is selected, the test tool waits for a trigger. After receiving a trigger, the waveform is displayed and the instruments is set to HOLD.

2

3

Open the Trigger Options menu.

Open the Trigger on Edge menu.

In most cases it is advised to use the Free Run mode:

4

 

 

 

Select Free Run, jump to Noise

 

 

 

 

reject Filter.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

5

 

 

 

Set Noise reject Filter to Off.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Observe that the key labels at the bottom of the screen have adapted to allow further selection of specific edge trigger settings:

49

Page 57
Image 57
Fluke 196C user manual Triggering on Edges, Open the Trigger on Edge menu, Select Free Run, jump to Noise, reject Filter