Appendices
Specifications A
Maximum Input
1000V dc or peak ac on any range
True RMS AC Voltage, AC-Coupled
Range
300mV
3V
30V
300V
750V
100mV
1000 mV
10V
100V
750V
Slow
—
—
—
—
—
1µ V
10µ V
100µ V
1mV
10mV
Resolution
Medium
10µ V 100µ V
1mV
10mV
100mV
—
—
—
—
—
Fast
100µ V
1mV
10mV
100mV
1V
—
—
—
—
—
Accuracy
| Linear Accuracy |
| ||
Frequency |
|
|
| |
Slow | Medium | Fast | ||
| ||||
|
|
|
| |
1 % + 100 | 1 % + 10 | 7 % + 2 | ||
50 | 0.2 % + 100 | 0.2 % + 10 | 0.5 % + 2 | |
0.5 % + 100 | 0.5 % + 10 | 0.5 % + 2 | ||
2 % + 200 | 2 % + 20 | 2 % + 3 | ||
5 % + 500 | 5 % + 50 | 5 % + 6 | ||
|
|
|
|
dB Accuracy
Slow/Med |
| Fast |
|
|
| ||
|
|
|
|
0.150.72
0.080.17
0.110.17
0.290.34
0.700.78
Power*
2 % + 10
0.4% + 10
1 % + 10
4 % + 20
10 % + 50
Max
Input at
Upper
Freq
750V
750V
750V
400V
200V
* Error in power mode will not exceed twice the linear accuracy specification
Accuracy specifications apply within the following limits, based on reading rate:
Slow Reading Rate: Between 15,000 and 99,999 counts (full range)
Medium Reading Rate: Between 1,500 and 30,000 counts (full range)
Fast Reading Rate: Between 150 and 3,000 counts (full range)
Decibel Resolution
Resolution
Slow & Medium
Fast
0.01 dB
0.1 dB