Fluke 45 user manual Specifications, Appendix A, Display Counts and Reading Rates, Response Times

Models: 45

1 122
Download 122 pages 22.93 Kb
Page 105
Image 105
Appendix A

Appendix A

Specifications

Introduction

Appendix A contains the specifications of the Fluke 45 Dual Display Multimeter. These specifications assume:

A 1-year calibration cycle

An operating temperature of 18 °C to 28 °C (64.4 °F to 82.4 °F)

Relative humidity not exceeding 90 % (non-condensing) (70 % for 1,000 krange

Accuracy is expressed as +(percentage of reading + digits).

Display Counts and Reading Rates

Rate

Readings per Second

 

 

Slow

2.5

Medium

5

Fast

20

 

 

* Ohms full range will typically be 98,000 counts

Full Range Display Counts

99,999*

30,000

3,000

RS-232 and IEEE-488 Reading Transfer Rates

Rate

Slow

Medium

Fast

Internal Trigger

Operation (TRIGGER 1)

2.5

4.5

4.5

Reading Per Second

Internal Trigger Operation

(TRIGGER 4)

1.5

2.4

3.8

Print Mode Operation

(Print set at 1)

2.5

5.0

13.5

Response Times

Refer to Section 4 for detailed information.

A-1

Page 105
Image 105
Fluke 45 user manual Specifications, Appendix A, Display Counts and Reading Rates, RS-232and IEEE-488Reading Transfer Rates