Fluke 45 Replacement Parts, •Quantity •Fluke Part Number •Part Description, Maintenance

Models: 45

1 122
Download 122 pages 22.93 Kb
Page 101
Image 101
Replacement Parts

Maintenance 6

Replacement Parts

Table 6-3. Performance Tests for mA Current Functions

Function

[(mA)

\(mA)

Range (Rate)

30mA

100mA

30mA

100mA

Input Level

+30 mA

+100 mA

30mA

100mA

Frequency

1kHz

1kHz

 

Display

 

 

 

Min

 

Max

 

 

 

29.982

30.018

99.93100.07

29.840

30.160

99.40100.60

Function

[

\

Table 6-4. Performance Tests for A Current Functions

 

 

 

 

 

 

 

 

Range (Rate)

Input Level

Frequency

 

Display

 

 

 

 

 

Min

 

Max

 

 

 

 

 

 

 

 

 

 

10 A

+10 A

9.975

 

10.025

10 A

10 A

1 kHz

9.890

 

10.110

 

 

 

 

 

 

Replacement Parts

Parts that can be replaced by the user are listed in Table 6-5. Other parts must be replaced by qualified service personnel only, following the procedures in the Service Manual.

Replacement parts and price information can be obtained from Fluke or an authorized Fluke distributor. To ensure prompt and efficient handling, include the following information with your order:

Quantity

Fluke Part Number

Part Description

Instrument Model and Serial number.

Table 6-5. Replacement Parts

ITEM

C40*

F1 W

F2 W

F3 W

F5 W RS40*

DESCRIPTION

Soft Carrying Case (Optional)

mA Input Fuse. F 500 mA, 250 V, (Fast blow), 1500 A breaking capacity

10 A Input Fuse. F 15 A, 250 V, (Fast blow), 10,000 A minimum breaking capacity; Or, F 15 A, 600 V, (Fast blow), 10,000 A minimum breaking capacity 820829

Line Fuse. T 125 mA, 250 V, (Slow blow)

See Supplement

RS-232 terminal interface cable. Connects Fluke 45 to any properly configured DTE connector (DB-25, male pins). IBM PC or IBM PC/XT, IBM PS/2 (Models 25,30,50, 60,70,80).

FLUKE PART NO.

838151

830802

822254

6-7

Page 101
Image 101
Fluke 45 user manual Replacement Parts, •Quantity •Fluke Part Number •Part Description, •Instrument Model and Serial number