Maintenance 6
Service
4.The following tests require a Fluke 5725A Amplifier (or equivalent) to be used with the 5700A. The input level for the performance test will be set on the 5700A but will be output from the 5725A Amplifier.
Connect a cable from the Output VA HI and LO connectors of the 5725A to the l0A and COM connectors on the Fluke 45. Select the function and range on the Fluke 45 and the input level from the 5700A using the values contained in Table
Service
If you suspect that the meter has failed, review this manual to make sure you are operating it correctly. If the meter still fails to operate properly, pack it securely (in its original container if available) and forward it, postage paid, to the nearest Fluke Service Center. Include a brief description of the problem. Fluke assumes NO responsibility for damage in transit.
To locate an authorized service center, call Fluke using any of the phone numbers listed below, or visit us on the World Wide Web: www.fluke.com
31 40 267 8200 in Europe
Table
Function
W
Range (Rate)
100 mV (slow)
1000 mV (slow)
300mV
3V
30V
300V
1000 V
—
—
300 mV
3V
30V
300V
750V
Input Level
Short +90 mV +900 mV short
300mV
+3 V
Short
15mV
300mV
3.0V 30 V 300 V 750 V
Frequency
—
—
—
—
—
—
—
—
—
—
—
1kHz
100kHz
1kHz
100kHz
1kHz
1kHz
1kHz
1kHz
| Display | |
|
|
|
Min |
| Max |
|
|
|
0.006 | ||
89.971 | 90.029 |
899.71900.29
299.90300.10
2.99903.0010
29.99030.010
299.90300.10
999.51000.5
0.00080.0008
(tone) | (tone) |
OL | OL |
— | <75 |
14.8715.13
13.7516.25
299.30300.70
284.50315.50
2.99303.0070
29.93030.070
299.30300.70
747.5752.5