Operating the Meter From the Front Panel | 3 |
Ranging |
Table
Fast Reading Rate | Medium Reading Rate | ||
|
|
|
|
Range | Full Scale | Range | Full Scale |
|
|
|
|
30 mA | 30.00 mA | 30 mA | 30.000 mA |
| |||
|
|
|
|
100 mA | 100.0 mA | 100 mA | 100.00 mA |
|
|
|
|
10 A | 10.00 A* | 10 A | 10.000 A* |
|
|
|
|
* 20 A for maximum of 30 seconds
Slow Reading Rate
Range | Full Scale |
|
|
10 mA | 9.9999 mA |
|
|
100 mA | 99.999 mA |
|
|
10 A | 9.9999 A |
|
|
Table
Fast Reading Rate
Range | Full Scale |
|
|
300 Ω | 300.0 Ω |
|
|
3 kΩ | 3.000 kΩ |
|
|
30 kΩ | 30.00 kΩ |
|
|
300 kΩ | 300.0 kΩ |
|
|
3 MΩ | 3.000 MΩ |
|
|
30 MΩ | 30.00 MΩ |
|
|
300 MΩ | 300 MΩ |
|
|
Medium Reading Rate
Range | Full Scale |
|
|
300 Ω | 300.00 Ω |
|
|
3 kΩ | 3.0000 kΩ |
|
|
30 kΩ | 30.000 kΩ |
|
|
300 kΩ | 300.00kΩ |
|
|
3 MΩ | 3.0000 MΩ |
|
|
30 MΩ | 30.000 MΩ |
|
|
300 MΩ | 300.0 MΩ |
|
|
Slow Reading Rate
Range | Full Scale* |
|
|
100 Ω | 98.000 Ω |
|
|
1000Ω | 980.00 Ω |
|
|
10 kΩ | 9.8000 kΩ |
|
|
100 kΩ | 98.000 kΩ |
|
|
1000 kΩ | 980.00 kΩ |
|
|
10 MΩ | 9.8000 MΩ |
|
|
100 MΩ | 98.0 MΩ ** |
|
|
*Typical
** Because of the method used to measure resistance, the 100 M (slow) and 300 M (medium and fast) ranges cannot measure below 3.125 M and 20 M respectively. "UL" (Underload) is shown on the display for resistances below these nominal points, and the computer interface outputs "+1
Table
Range
Full Scale
1000 Hz | 999.99 Hz |
|
|
10 kHz | 9.9999 kHz |
|
|
100 kHz | 99 999 kHz |
|
|
1000 kHz | 999.99 kHz |
|
|
1 MHz | 9.9999 MHz |
|
|
* Fast (F) reading rate has one digit of resolution less.