Function Description

Functional Description 4

Section 4

Function Description

4.1 Functional Description

High-Range Containment Monitor Detector 877-1

The high-range containment monitor detector is an ion chamber detector that has the appearance of a six-inch diameter domed cylinder about seven inches long, mounted on an L shaped bracket. Inside the cylinder are two flange-mounted electrodes consisting of 31 flat, disk-shaped plates, each about four inches in diameter, stacked, and mounted on disk rods. The 31 disks form two groups, interleaved with each other, 16 collection disks and 15 signal disks. Because of the interleaving, they appear as only one stack, but the collection disks are mounted on three collection disk-posts and the signal disks on three signal disk-posts. Spacers on the posts keep the disks separated so they do not come in contact, and clearance holes in the disks allow posts of the opposite polarity to pass through without contact. The assembly has the appearance of a large air capacitor. The mounting posts are attached to the mounting flange through insulating spacers, so the flange and housing will be neutral with respect to the charges applied to the electrodes. The collecting diskposts are elongated beyond the last collecting disk to support a cup-shaped liner having the same potential as the collecting disks, thus becoming part of the collector.

The whole assembly is covered by the six-inch diameter housing which contacts only the neutral mounting flange. The mounting flange is pierced by three holes. One hole supports the exhaust tube used for exhausting and back filling the chamber. The other two support two 2-pin connectors, one for each electrode. One pin in each pair is connected to the neutral mounting flange. When the coaxial signal cable is connected to this connector, the cable shield is connected to the neutral pin.

The entire chamber is filled with a mixture of helium (2%) and nitrogen at atmospheric pressure and sealed.

CAUTION

The seal on this chamber must not be broken. To do so would alter the calibration and specified energy response of the system.

4.2 Readout Module 876A-1

NOTE

Drawings 876A-1-3H and 876-1-3A serve as interconnecting diagrams for tracing signals between printed circuit boards. In addition, drawing 876-1-3A contains the main power supply