Introduction
General Description 1
Section 1
Introduction
1.1 General Description
Containment Monitor 875
NOTE
The 875 Containment Monitor has been qualified for use in Nuclear Safety related applications. Qualification Test Reports, 950.301, 950.308A, and 950.310A define the parameters that have been verified by test and are available for purchase.
Containment Monitor 875 functions as an accident monitor for reactor containments, refer to Figure
1.2 Specifications
Detector 877-1
Radiation Detected | Photons above 60 keV |
Range | 1 R/h to 1E7 R/h |
Energy Response | ± 20% from 80 keV to 2 MeV |
Chamber Construction | Stacked parallel plate, 316 stainless steel housing, 3 |
| terminal, guarded ion chamber |
Saturation | >80% up to full scale on instrument |
Collection Voltage | 450 to 550 VDC |
Current Output | |
Chamber Capacitance | 435 picofarads (nominal) |
Voltage Withstand and Leakage | |
| connector pins |
Capacitance | Approximately 430 pf |