Maintenance, Calibration, and Troubleshooting

5

Operation of Latch Circuit

U202, pins 1, 5, L (Enable pulse active; i.e. low)

From these conditions, the following must hold true.

Pin

Comment

4

Depends on 6

3

Depends on 2

Further analysis is contained in the truth table (Table 5-2).

Figure 5-2.Latch Circuit

Table 5-1. Truth Table for Pass and Fail Conditions in The Latch Circuit

(Circuit assumed initially in Pass Condition)

 

Fail Condition

 

Pass Condition

 

Pin

Voltage

Voltage Comments

Pin

State

Comments

State

 

 

 

 

 

4

L (low)

Since 6 is H (high)

4

H

Because 6 is L (also 5)

3

H (high)

Since 1 & 2 are Low

3

L

Because 2 is H

13

H

Connected to 3

8

H

Connected to 4

11

L

Since 13 is H

13

L

Connected to 3

9

L

Connected to 11

10

L

Because 8 is H

5-11

Page 41
Image 41