Victoreen 875
Operators Manual
Hermetic Integrity | ||
| atmosphere | |
Chamber Fill Gas | Nitrogen/helium (2%) at atmospheric pressure | |
Dimensions (W x H x D) with Mounting Bracket | 9.25 x 12.5 x 10.06 in | |
| (234.95 x 317.5 x 255.52 mm) | |
Weight | Approximately 18 lb (8.17 kg) | |
Temperature, Storage | 40° to 160°F (4° to 71°C) | |
Temperature, Operating | 40° to 160°F (4° to 71°C) | |
| 375°F (180°C) peak, accident, 3 hours | |
Relative Humidity | 100% (waterproof) | |
Total Integrated Dose | 2E8 Rads + 10% margin, 60Co | |
Chemical Spray | 0.45 Gallons/minute/ft2 (0.28 molar H3BO3, 3000 ppm | |
| Boron, and NaOH; ph of 11, 24 hrs | |
LOCA Test Duration | 28 days |
|
(Consult Fluke Biomedical, Radiation Management Services, for a complete description of Qualification Test Results)
Readout Module 876A-1
Range
Input Current
Recorder Output optional)
Computer Output optional)
System Accuracy (during "all" conditions)
Power Requirements:
maximum
Temperature Coefficient
Dimensions (W x H x D)
Weight
Environmental Parameters
1 to 107 R/h
Minimum: 6.5 to 7.5 x
Maximum: 6.5 to 7.5 x
0 to +1 VDC, logarithmic (0 - 10 mV and 0 – 5 V
0 to +5 VDC, logarithmic (0 - 100 mV and 0 - 50 mV
(Other Outputs Available Upon Request)
Accumulative @ Meter +36% of input radiation Analog Outputs +28% of input radiation
a)AC Voltage: 108 to 132 VAC, RMS @ 60 ± 3 Hz
b)Battery Power: 22 to 32 VDC @ 600 mA DC
0.40%/°C, + 0.25 R/h/°C
8.46 x 5.25 x 15.21 in
(214.88 x 133.35 x 386.33 mm)
20 lbs (9.07 kg)
Temperature (Storage): 40° to 140°F (4° to 60°C)