Fluke T100/120/140

Users Manual

Voltage Test with RCD Trip Test

(only T140 VDE)

1)Press Button on rear side - measurement point light (10).

2)Measure between L and PE.

3)The RCD is tripping

The RCD could be so long tripped like the measuring point lighting is active and L is measured against PE. After a voltage test wait approx.. 50 seconds to trip the RCD again.

Restistance Test (only T140 / T140 VDE)

Make sure that UUT is not live.

Check that UUT is not live by carrying out a double-pole voltage test.

Connect both test probes with UUT. Press Button on rear side - for measurement point light (10) and read value on the display.

The resistance range is 1…1999 Ω at a resolution of 1 Ω.

The resistance test is active for 20 se- conds after having pressed the button on the rear.

If during the resistance measurement a voltage is present the instrument switch automatically to voltage measurement.

Continuity Test / Diode Test

Make sure that UUT is not live. Test volta- ge polarity at handle test probe is positive

(+).

Check that UUT is not live by carrying out a double-pole voltage test.

Connect both test probes with UUT. A signal sound is audible for continuity and the LED for continuity Rx/Ω is illuminated.