PROBLEM

PROBABLE CAUSE

CORRECTIVE MEASURES

4) Unusually high body fat readings

a) Too much light present during the

a) Do not lift the Light Wand from the

(40-50%)

reading.

subject's biceps until a "2" appears

 

 

on the screen during the first

 

 

measurement or the printer starts

 

 

printing after the second

 

 

measurement.

 

 

b) Use the Light Shield to block the

 

 

external light. Always avoid

 

 

measurements under strong lights

 

 

(e.g. direct sunlight) because errors

 

 

may occur.

5) Printer Stops

a) Low batteries.

a) Replace batteries (see page 3).

 

b) Bad battery contact.

b) Remove batteries and clean tabs

 

 

with isopropyl alcohol (rubbing

 

c) Bad AC connection.

alcohol).

 

 

c) Unplug the AC Adapter from the

 

 

wall socket and from the unit. Clean

 

 

the adapter contact with isopropyl

 

 

alcohol (rubbing alcohol).