NOTE - IT IS IMPORTANT TO:

!Aim the silver stripe of the Light Wand vertically toward the upper portion of the body.

!Keep the body site to be measured relaxed (do not permit subject to flex muscle area being measured).

!Use the Light Shield to block the external light. ALWAYS avoid measurements under strong lights (e.g. direct sunlight) because errors may occur.

!Press ENTER.

The display will then show a decimal point following the number "1". The number "2" will then be displayed.

CAUTION: Do not move the Light Wand while the decimal point is illuminated. The actual measurement is being made at that time.

!

Remove the Light Wand and replace it on the same site on the body.

!Press ENTER.

The display will show a decimal point following the number "2".

6)The Relative Body Fat Reading is Displayed

NOTE: This relative body fat is not a direct "Percentage" reading. This value, however, can be used as a guide to monitor changes in fat distribution within the location measured.

To Measure Another Site or Repeat a Measurement

Place the Light Wand firmly on the particular body site and follow steps outlined in 6 and 7.

OFF Message

An OFF Message will appear and the FUTREX-5000A/WL Automatic Power Down feature will turn the unit OFF if the user has not pressed any key for 6 minutes. To use the unit again, turn the Power Switch OFF and then ON again.