Connecting Field Wiring

 

Leads for the individual channels are connected through the channel

 

terminals on the terminal modules. Multimeters and signal generators can

 

be connected to the direct terminals or tree terminals on the terminal

 

modules. Figure 2-9 shows the terminal module channel terminals for the

 

16-Channel FET Multiplexer Module (HP E1351A) and the 16-Channel

 

Thermocouple FET Multiplexer Module (HP E1353A).

Wiring Guidelines

If possible, use shielded cables with the shields connected to the Guard

 

(G) terminals and to the low connection near the measurement point.

 

Be sure the wires make good connections on the screw terminals.

 

For thermocouples, connect the Guard terminal to the

 

thermocouple’s shield lead and the low connection near the

 

measurement point.