Scanning a Switchbox without a Downloaded Scan List

This example shows a FET switchbox connected to a multimeter with the analog bus cable. The switchbox and multimeter are separate VXIbus instruments. The multimeter has a secondary address of 03, and the switchbox has a secondary address of 04. The multiplexer is triggered by the *TRG command. OUTPut ON is enabled, and the multimeter is triggered by the OUTPut ON trigger to its "External Trigger" port. The following program illustrates the procedures:

10

DIM Rdgs(1:16)

!Dimension array to store readings

20

OUTPUT 70903;"*RST;*OPC?"

!Clear the multimeter; *OPC?

 

 

ensures reset is completed before

 

 

program continues

30

ENTER 70903;A

 

40

OUTPUT 70903;"*CLS"

 

50

OUTPUT 70904;"*RST"

!Reset multiplexer

60

OUTPUT 70904;"*CLS"

 

70

OUTPUT 70903;"CONF:VOLT:DC 58.1"!Configure the multimeter

80

OUTPUT 70903;"TRIG:SOUR EXT"

!External trigger source

90

OUTPUT 70903;"TRIG:COUN 16"

!Set for 16 triggers

100

OUTPUT 70903;"INIT"

!Initialize multimeter wait for

 

 

trigger

110

OUTPUT 70904;"TRIG:SOUR BUS" !Trigger on *TRG command

120

OUTPUT 70904;"SCAN:MODE VOLT"!Configure for voltage

130

OUTPUT 70904;"SCAN:PORT ABUS"!Enables analog bus

140

OUTPUT 70904;"SCAN (@100:115)"!Enter scan list

150

OUTPUT 70904;"OUTP ON"

!Enable mainframe trig out port

160

OUTPUT 70904;"INIT"

!Close first channel

170

FOR I = 1 TO 16

!16 channels

180

OUTPUT 70904;"*TRG"

!Trigger for channel advance

190

WAIT .25

!Allow time for measurement

 

 

before next advance

200

NEXT I

!Increment counter

210

OUTPUT 70903;"FETC?"

!Multimeter retrieves readings

 

 

from mainframe memory when

 

 

scan completes

220

ENTER 70903;Rdgs(*)

!Put readings in array

230

PRINT Rdgs(*)

!Print

240

END

 