Scanning a Range of Switchbox Channels
You can scan a range of channels of a switchbox consisting of single or multiple multiplexer modules. Scanning involves sequentially closing each channel on a range of specified channels. The selected TRIGger mode determines when the channel closure will advance. During scanning, the FET which was previously closed opens before the next FET closes.
You can use any of the TRIGger modes to advance the channel list. TRIGger:SOURce EXTernal requires an external trigger source to be connected to the HP E1300/1301 Mainframe "Event In" port. For computer controlled or
For the fastest scan rate (up to 100 K switches/sec.) use TRIGger:SOURce DBUS. Use a custom cable (see “ Di gital Bus and Triggering” on page 15) to connect the multimeter handshake lines (Voltmeter Complete and External Trigger) to the multiplexer digital bus.
Figures
∙A Tree Terminals of each terminal module connect to each other and to the multimeter in Figure
∙
∙HP E1300A/E1301A Mainframe’s "Trig Out" pulse synchronizes the switchbox with the multimeter.
∙Multimeter
∙Switchbox
∙Computer is an HP Series 200/300 Computer with HP BASIC using
Enter and Execute:
10 | OUTPUT 722;"TRIG EXT;DC 10" | !Sets multimeter to external |
|
| triggers and to measure dc volts |
20 | OUTPUT 70914;"OUTP ON" | !Enables "Trig Out" port |
30OUTPUT 70914;"TRIG:SOUR BUS" !Sets switchbox to receive bus triggers
40OUTPUT 70914;"SCAN:MODE VOLT"!Setup switchbox to measure voltage
50OUTPUT 70914;"SCAN:PORT ABUS"!Closes the appropriate tree isolation switches while scanning; automatically makes connection to the analog bus and tree terminals.
36 Using the HP E1351A/53A FET Multiplexer Modules | Chapter 3 |