ABORt

The ABORt subsystem stops a scan in progress when the scan is enabled via the interface, and the trigger modes are TRIGger:SOURce BUS or TRIGger:SOURce HOLD.

Subsystem Syntax

ABORt

Comments ABORt Operation. Resets the trigger system and places triggers in idle. The scan is stopped without resetting the scan conditions. The scan conditions stay in the current configuration.

Stopping Scans Enabled from Interface. When a scan is enabled from the interface, use an interface CLEAR command or the HP E1301 front panel "Reset Instr" or "Clear Instr" key to stop the scan.

When the scan is enabled from the interface and the trigger source is

TRIGger:SOURce BUS or TRIGger:SOURce HOLD, use ABORt or the HP E1301 front panel "Reset Instr "or "Clear Instr "keys to stop the scan.

Stopping Scans Enabled from Front Panel. When a scan is enabled from the HP E1301 front panel execute *RST over the interface or use the front panel "Reset Instr "or "Clear Instr" keys to stop the scan.

Related Commands. ARM, INITiate:CONTinuous, [ROUTe:]SCAN, TRIGger

Example Stopping a Scan with ABORt

TRIG:SOUR BUS

Bus is trigger source.

INIT:CONT ON

Set continuous scanning.

SCAN (@100:115)

Sets channel list.

INIT

Starts scanning cycle.

.

 

.

 

.

 

ABOR

Aborts scan in progress.

56 HP E1351A/53A 16-Channel FET Multiplexer Command Reference