OUTPut
The OUTPut subsystem enables or disables the "Trig Out" port of the HP
E1300/E1301 Mainframe.
Subsystem Syntax | OUTPut |
| [:STATe] <state> |
| [:STATe]? |
[:STATe] | OUTPut[:STATe] <state> enables/disables the "Trig Out" port on the rear |
| panel of the HP E1300/E1301 Mainframe. OUTPut[:STATe] ON 1 enables |
| the port and OUTPut[:STATe] OFF 0 disables the port. |
Parameters
| Parameter Name | Parameter Type | Range of Values |
|
|
|
|
| state | Boolean | 0 1 ON OFF |
|
|
|
|
Comments | ∙ Downloaded Scans: The OUTPut subsystem does not work with |
downloaded scan lists. The scan list is downloaded automatically if you use TRIG:SOUR DBUS or TRIG:SOUR IMM (FET multiplexers only). OUTPut works with all other scanning methods.
∙Enabling "Trig Out" Port: When enabled, the "Trig Out" port is pulsed each time a channel is closed during scanning. When disabled, the "Trig Out" port is not pulsed.
∙Output Pulse: The pulse is a +5 V
∙"Trig Out" Port Reserved by Switchboxes: The "Trig Out" resource is a system resource, and can only be reserved by one instrument at a time.
∙Related Commands: [ROUTe:]SCAN, TRIGger:SOURce
∙*RST Condition: OUTPut[:STATe] OFF (port disabled)
Example Enabling "Trig Out" Port
OUTP ON | Enables "Trig Out" port for pulse |
| output. |
[:STATe]? OUTPut[:STATe]? queries the present state of the "Trig Out" port. The command returns 1 if the port is enabled, or 0 if disabled.
Example Query "Trig Out" Port State
OUTP ON | Enables "Trig Out" port for pulse |
| output. |
OUTP:STAT? | Query port enable state. |
62 HP E1351A/53A | Chapter 5 |