MultimeterCommand andParameter Opcodes

Table B-1 contains the multimeter command and parameter opcodes. The opcodes used to query the parameter settings are also included.

Table B-1. Multimeter Command and Parameter Opcodes

Multimeter

Command

Parameter

Value

 

Query

 

Query

Parameter

Opcode

Opcode

 

 

Opcode

 

Response

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Measurement

04

00

DCV

 

05

 

parameter

Function

 

01

ACV

 

 

 

opcode

 

 

02

(4-wire) OHMS

 

 

 

 

Range

02

00

125 mV / 256 Ω

 

03

 

parameter

 

 

01

1V / 2.048 kΩ

 

 

 

opcode

 

 

02

8V / 16.384 kΩ

 

 

 

 

 

 

03

64V / 131 kΩ

 

 

 

 

 

 

04

300V / 1.048 MΩ

 

 

 

 

 

 

05

Autorange

 

 

 

 

 

 

 

 

 

 

 

 

Aperture Time

00

00

267 ms

 

01

 

parameter

 

 

01

16.7 ms

 

 

 

opcode

 

 

02

320 ms

 

 

 

 

 

 

03

20 ms

 

 

 

 

 

 

04

2.5 ms

 

 

 

 

 

 

05

100 μs

 

 

 

 

 

 

06

10 μs

 

 

 

 

 

 

 

 

 

 

 

 

Autozero

08

00

On

 

09

 

parameter

 

 

01

Off

 

 

 

opcode

 

 

02

Once

 

 

 

 

 

 

 

 

 

 

 

 

Offset

36

00

Off

 

37

 

parameter

Compensation

 

01

On

 

 

 

opcode

 

 

 

 

 

 

 

 

Trigger Source

06

00

Immediate

 

29

 

parameter

 

 

01

External

 

 

 

opcode

 

 

02

Hold

 

 

 

 

 

 

03 - 0A

TTL trigger lines

 

 

 

 

 

 

 

0 - 7

 

 

 

 

 

 

 

 

 

 

 

 

Trigger Count

39

<upper byte>

1 - 16,777,215

 

33

 

24-bit unsigned

 

 

<middle byte>

(1),(3)

 

 

 

number

 

 

<lower byte>

 

 

 

 

(2)

 

 

 

 

 

 

 

 

Trigger Delay

23

<upper byte>

0 - 16.777215s

 

31

 

24-bit unsigned

 

 

<middle byte>

(1)

 

 

 

number

 

 

<lower byte>

 

 

 

 

(2)

 

 

 

 

 

 

 

 

Voltmeter

40

8-bit binary number

0 - 7

 

41

 

8-bit binary number

Complete Signal

 

(7)

(TTL trigger lines)

 

 

 

 

Destination

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Sample Count

38

<upper byte>

1 - 16,777,215

 

32

 

24-bit unsigned

 

 

<middle byte>

(1),(4)

 

 

 

number

 

 

<lower byte>

 

 

 

 

(2)

 

 

 

 

 

 

 

 

Sample Source

07

00

Immediate

 

30

 

parameter

 

 

01

Timer

 

 

 

opcode

 

 

02

Software

 

 

 

 

 

 

03

Control Register

 

 

 

 

 

 

 

 

 

 

 

Sample Period

10

<upper byte>

76μs - 65.534ms

11

 

16-bit unsigned

(5)

 

<lower byte>

(1)

 

 

 

number

 

 

 

 

 

 

 

(2)

 

 

 

 

 

 

 

 

Software Sample

35

—-

—-

—-

...

 

(6)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Trigger Arm

12

00

Un-armed

 

—-

 

—-

 

 

01

Armed

 

 

 

 

 

 

 

 

 

 

 

Trigger Immediate

13

—-

—-

—-

...

 

 

 

 

 

 

 

 

Send Error

15

—-

—-

—-

—-

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Table B-1 Continued on Next Page

96 HP E1351A/53A Register-Based Programming

Appendix B