SETTling[:TIME] [ROUTe:]SETTling[:TIME] <channel_list> sets the delay between receipt of a channel closing trigger, and the pulse when the FET channel is closed. You must specify one and only one channel for each card in the instrument which has a channel in the channel list, and the <time> will then apply to all channels on the card. The range is 1 to 32768 μsec.

Parameters

Parameter Name

Parameter Type

Range of Values

 

 

 

time

Numeric

1E-6 32768E-6 MIN MAX

 

 

 

channel_list

Numeric

only one channel

 

 

 

Comments Digital Bus Handshaking: This command allows a user-defined settling time to allow a measurement to be made before the next channel is closed.

Related Commands: TRIGger:SOURce DBUS, TRIGger:SOURce IMMediate

Example Setting 16 μsec SETTling TIME

TRIGger:SOUR DBUS

Triggering on the digital bus

 

handshake lines

SETTling:TIME 16E-6

Delay = l6μsec

SETTling[:TIME]? [ROUTe:]SETTling[:TIME]? [MIN MAX] <channel_list> returns the

current setting for a specified channel. MIN or MAX returns the minimum and maximum allowable settings.

Parameters

Parameter Name

Parameter Type

Range of Values

 

 

 

MIN MAX

Numeric

MIN = 1 μsec, MAX = 32768 μsec

 

 

 

channel_list

Numeric

only one channel

 

 

 

Comments Returned value is given in exponential notation.

Example Set and Query the SETTling TIME

SETT:TIME 16E-6

Sets delay = 16μsec.

SETT:TIME?

Returns + 1.600000E-005.

Chapter 5

HP E1351A/53A 16-Channel FET Multiplexer Command Reference 69