Command Quick Reference

The following tables summarize SCPI and IEEE 488.2 Common (*)

Commands for the 16-Channel FET Multiplexer Modules.

SCPI Commands Quick Reference

 

Command

Description

 

 

 

 

 

 

ABORt

 

Abort a scan in progress.

 

 

 

ARM

:COUNt <number> MIN MAX

Multiple scans per INIT command.

 

:COUNt? [MIN MAX]

Query number of scans.

 

 

 

DISPlay

:MONitor:CARD <number> AUTO

Selects module to be monitored.

 

:MONitor[:STATe] <mode>

Selects monitor mode.

 

 

 

INITiate

:CONTinuous <mode>

Enables/disables continuous scanning.

 

:CONTinuous?

Query mode.

 

[:IMMediate]

Starts a scanning cycle.

 

 

 

OUTPut

[:STATe] <state>

Enables/disables "Trig Out" pulse.

 

[:STATe]?

Query state of trig out pulse.

 

 

 

[ROUTe:]

CLOSe <channel _list>

Close channel(s).

 

CLOSe? <channel _list>

Query channel(s) closed.

 

OPEN <channel _list>

Open channel(s).

 

OPEN? <channel _list>

Query channel(s) opened.

 

SCAN <channel _list>

Define channels for scanning.

 

SCAN:MODE NONE VOLT RES FRES

Selects appropriate tree Isolation switches for measurements.

 

SCAN:MODE?

Query state of scan mode.

 

SCAN:PORT NONE ABUS

Enables tree isolation switches closure as part of a

 

SCAN:PORT?

measurement.

 

Query analog bus state.

 

SETTling[:TIME] <channel_list>

Sets delay between trigger pulse and channel closed pulse.

 

SETTling[:TIME]? [MIN MAX] <channel_list>

Returns current parameter in SETTling[:TIME].

 

 

 

STATus

:OPERation:ENABle

Enables the Operation Status Register to set a bit in the

 

 

Status Register.

 

:OPERation[:EVENt]?

Returns status of Operation Status Register.

 

 

 

SYSTem

:CDEScription? <number>

Returns description of module in a switchbox.

 

:CPON <number> ALL

Sets specified module to its power-on state.

 

:CTYPe? <number>

Returns the module type.

 

:ERRor?

Returns error number/message in a switchbox error queue.

 

 

 

TRIGger

[:IMMediate]

Causes a trigger to occur.

 

:SOURce BUS

Trigger source is *TRG.

 

:SOURce DBUS

Trigger source is Digital Bus.

 

:SOURce EXTernal

Trigger source is "Event In" port.

 

:SOURce HOLD

Hold off triggering.

 

:SOURce IMMediate

Continuous (internal) triggers.

 

:SOURce?

Query scan trigger source.

 

 

 

 

 

 

 

 

IEEE 488.2 Common Commands Quick Reference

Command

Title

 

Description

 

 

 

 

 

 

 

 

*RST

Reset

 

Opens all channels and invalidates current channel list for scanning.

 

 

 

Sets ARM:COUN 1, TRIG:SOUR IMM, and INIT:CONT OFF.

 

 

 

 

*TRG

Bus Trigger

 

When scan is enabled and trigger source is TRIG:SOUR BUS, use the *TRG

 

 

 

command to trigger the switchbox to advance the scan.

 

 

 

 

*TST?

Self-Test

 

Power on interrupt returns +110.

 

 

 

Incorrect MFR ID or model code returns +103.

 

 

 

Test pass returns 0.

 

 

 

 

 

 

 

 

Chapter 5

HP E1351A/53A 16-Channel FET Multiplexer Command Reference 77