Installation and Setup
Power Requirements
Power Requirements
The only physical installation of your Agilent spectrum analyzer is a connection to a power source.
Line voltage does not need to be selected.
WARNING Failure to ground the analyzer properly can result in personal injury. Before turning on the analyzer, you must connect its protective earth terminals to the protective conductor of the main power cable. Insert the main power cable plug into a socket outlet that has a protective earth contact only. DO NOT defeat the
If you are using an autotransformer, make sure its common terminal is connected to the protective earth contact of the power source outlet socket.
This is a Safety Class 1 Product (provided with a protective earthing ground incorporated in the power cord). The mains plug shall only be inserted in a socket outlet provided with a protective earth contact. Any interruption of the protective conductor inside or outside of the product is likely to make the product dangerous. Intentional interruption is prohibited.
CAUTION VENTILATION REQUIREMENTS: When installing the product in a cabinet, the convection into and out of the product must not be restricted. The ambient temperature (outside the cabinet) must be less than the maximum operating temperature of the product by 4° C for every 100 watts dissipated in the cabinet. If the total power dissipated in the cabinet is greater than 800 watts, then forced convection must be used.
This analyzer has autoranging line voltage input. Be sure the supply voltage is within the specified range.
NOTE | For more information regarding analyzer specifications, see the Specifications |