| Options and Accessories | |
|
| Options | |
|
|
| |
Name | Option | Description | |
Number | |||
|
| ||
|
|
| |
Phase Noise | 226 | Option 226 is an automated phase noise (dBc/Hz) | |
Measurement |
| measurement solution, providing the ability to measure | |
Personality |
| SSB noise, RMS noise, DANL, as well as the | |
|
| generation of a phase noise vs. log offset frequency, and | |
|
| phase noise vs. time. High Stability Frequency | |
|
| Reference (Option 1D5) is recommended. Narrow | |
|
| resolution bandwidth (Option 1DR) is highly | |
|
| recommended. Without Option 1DR, the minimum offset | |
|
| frequency will be 10 kHz. | |
|
|
| |
1DS | The | ||
|
| (lowers the noise floor) by approximately 16 dB. | |
|
|
| |
Rack Mount Kit with | 1CP | Option 1CP provides the parts necessary to mount the | |
Handles |
| analyzer in a standard 19 inch (482.6 mm) equipment | |
|
| rack. It includes front handles and rack slides for added | |
|
| convenience. Rack mount height is 8.75 in (222.3 mm). | |
|
|
| |
RF and Digital | B74 | Option B74 includes the RF Communications Hardware | |
Communications |
| (Option B7E), Digital Signal Processing and Fast ADC | |
Hardware |
| (Option B7D), | |
|
| 1D6), Memory Extension (Option B72), High Stability | |
|
| Frequency Reference (Option 1D5), Preamplifier (Option | |
|
| 1DS), and Narrow Resolution Bandwidth (Option 1DR) | |
|
| assemblies. These options are required for optimum | |
|
| performance of the cdmaOne (Option BAC) or GSM | |
|
| measurement personality (Option BAH). | |
|
|
| |
RF Communications | B7E | Option B7E provides the communications hardware | |
Hardware |
| required for many digital communication | |
|
| measurements. Option B7E must be ordered with | |
|
| Option B7D (digital signal processing and fast ADC) and | |
|
| Option 1D5 (high stability frequency reference). This | |
|
| provides burst carrier trigger capability. | |
|
|
| |
1AX | Option 1AX allows you to control your analyzer from a | ||
Interface |
| computer that uses an | |
|
| meter | |
|
| (Agilent | |
|
| printers, a | |
|
| software, and the programmer’s guide for your analyzer. | |
|
| This program allows you to download spectrum analyzer | |
|
| display or data files to a personal computer using | |
|
| Microsoft Word or Microsoft Excel. IntuiLink Toolbar | |
|
| installation instructions are included with the | |
|
| Option 1AX allows the analyzer to copy its display to a | |
|
| printer. | |
|
|
|
Chapter 5 | 81 |