Front and Rear Panel Features
| PRESEL TUNE OUTPUT provides a signal to control external preselected mixers if |
| External Mixing (Option AYZ) is installed. |
11 | Card Slot Identification Numbers. Refer to Table |
| card slot versus option card compatibility information. |
12 | 10 MHz REF IN accepts an external frequency source to provide the 10 MHz, − 15 |
| to +10 dBm as a timebase. |
|
|
NOTE | It is not necessary to connect the 10 MHz REF OUT to the 10 MHz REF IN on the |
| rear panel of the analyzer. Doing so will result in a Frequency |
| Reference Error message. |
|
|
13 | 10 MHz REF OUT provides a 10 MHz, 0 dBm minimum, timebase reference |
| signal. |
14 | Power On Selection selects an analyzer power preference. |
The PWR ALWAYS ON setting turns the analyzer on whenever external power is applied. This mode is useful if an external power switch is used to control a rack of several instruments. Nevertheless, if you set the analyzer to standby using the front panel Standby key (see Figure
23)and the external power is removed and restored within 20 seconds, the analyzer will remain in standby.
Power Always OnAnalyzer state before removing | A lapse in power | A lapse in power |
power | < 20 sec. | > 20 sec. |
|
|
|
On | On | On |
|
|
|
Standby | Standby | On |
|
|
|
The PWR NORM setting assigns analyzer on/off control to the
Analyzer state before removing | A lapse in power | A lapse in power |
power | < 20 sec. | > 20 sec. |
|
|
|
On | On | Standby |
|
|
|
Standby | Standby | Standby |
|
|
|