Section V

 

 

 

Model 3555B

 

Table 5-6. Front Panel Trouble Analysis

 

 

 

 

 

 

 

 

3555B SHOULD INDICATE

SET ACTUALLY

CORRECTIVE

INPUT CONDITIONS

FUNCTION

(RANGE + METER)*

INDICATES

ACTION

FUNCTION: VF/Nm

DIAL BAT

In green area, BAT GOOD

 

Replace battery

1kHz, 0dBm, 600 BAL

Input: TMS, TERM

 

 

 

 

VF/Nm: 600 BAL

0dBm ±0.2dBm

 

Refer to Paragraph 5-34

 

Change INPUT to BRDG

 

 

 

 

RANGE to +10dBm

+6dBm ±0.2dBm

 

Refer to Paragraph 5-34

 

Depress 900 BAL

+4.2dBm ±0.2dBm

 

Refer to Paragraph 5-34

 

INPUT to TERM

 

 

 

 

RANGE to 0dBm

-0.1 5dBm ±0.2dBm

 

Refer to Paragraph 5-34

FILTERS

INPUT: NOISE TERM

0dBm

 

Refer to Paragraph 5-37

 

NOISE WTG: 3kHz FLAT

 

 

 

 

VF/Nm, 600 BAL

0dBm ±0.2dBm

 

Refer to Paragraph 5-37

 

Change to C MSG

0dBm +±0.2dBm

 

Refer to Paragraph 5-37

 

Change to 15kHz FLAT

0dBrn ±0.2dBm

 

Refer to Paragraph 5-37

 

Change to PROG

0dBm ±1dBm

 

Refer to Paragraph 5-37

FUNCTION: CARRIER

 

 

 

 

20kHz, 0dBm

INPUT: TMS, TERM

 

 

 

600 BAL

FUNCTION: CARRIER

0dBm

 

Refer to Paragraph 5-34

 

600 BAL

 

 

 

 

Change INPUT to BRDG

 

 

 

 

RANGE to +10dBm

+6dBm ±0.5dBm

 

Refer to Paragraph 5-34

 

Depress 135 BAL

 

 

 

 

RANGE to +20dBm

+12.6dBm ±0.5dBm

 

Refer to Paragraph 5-34

 

Change INPUT to TERM

 

 

 

 

RANGE to 0dBm

-2.2dBm ±0.5dBm

 

Refer to Paragraph 5-34

Change to 75

UNBAL INPUT: TMS, TERM

 

 

 

 

FUNCTION: 75 UNBAL

0dBm ±0.2dBm

 

Refer to Paragraph 5-34

 

RANGE: 0dBm

 

 

 

 

Change INPUT to BRDG

 

 

 

 

RANGE to 10dBm

±6dBm +0.2dBm

 

Refer to Table 5-8

 

Change INPUT back to

 

 

 

 

TERM

 

 

 

 

RANGE to 0dBm

0dBm ±0.2dBm

 

Refer to Table 5-8

RANGE

 

 

 

 

1kHz, 600 BAL,

 

 

 

 

LEVEL +10dBm

Change RANGE to +10dBm

 

 

 

 

FUNCTION: VF/Nm 600 BAL

+10dBm ±0.2dBm

 

See Paragraph 5-35

LEVEL -10dBm

Change RANGE to -10dBm

-10dBm ±0.2dBm

 

See Paragraph 5-35

LEVEL -20dBm

Change RANGE to -20dBm

-20dBm ±0.2dBm

 

See Paragraph 5-35

5-12

Page 46