Hughes & Kettner TrilogyTM manual Troubleshooting, Trilogy Manual, english

Models: TrilogyTM

1 56
Download 56 pages 30.26 Kb
Page 11
Image 11
6 Troubleshooting

Trilogy™ Manual 1.1

6 Troubleshooting

Trilogy™ won’t power up when you switch it on.

It‘s not getting AC power. Check the mains cord to see if it is connected and firmly seated.

The mains fuse is defective. Ensure it is replaced with another fuse of the same rating.

Trilogy™ is connected properly, but no sound is audible.

The amp is set to STANDBY.

The GAIN or MASTER knob is turned all the way down.

The effects loop is active.

The anode fuse has blown. Before having it replaced with a fuse of the same rating, have the power amp tubes checked for defects.

The fuse for the tube heating tripped (the tubes don’t glow). Ensure that it is replaced with a fuse bearing the same rating. Always have a service technician replace fuses.

The sound is washed out or muddy when you switch an effects processor on.

The signal processor provides a wet signal that is blended with the dry or original signal. Depending on the type of effect, the processor may be returning a dry signal back along with wet signal, which causes phase cancellations when mixed to the dry signal in Trilogy™’s parallel loop. To prevent phase cancellations, switch the effects loop to SERIAL or turn the dry signal all the way down on the signal processor.

The amp starts making ringing or hissing noises.

One or several tubes are microphonic. Replace the defective tube with another of the same type.

Signs of tube wear such as increased microphonics and noise, treble loss, weak power output or muddy sound begin reappearing just a few hours after replacing tubes.

The wrong tubes were installed when old tubes were replaced or the amp was not biased properly. Take the amp to a professional to correct the problem.

english

11

Page 11
Image 11
Hughes & Kettner TrilogyTM manual Troubleshooting, Trilogy Manual, english