Transmitted-light stand, bright field, 20 W
Follow the safety instructions given on page 7.
Fit the
Fitting the bulb
Never touch the bulb when it is hot. You may sustain burns.
•Do not touch the new 6 V/20 W halogen bulb directly with your fingers, because finger marks can cause it to crack when heated.
Remove finger marks immediately with a cloth dipped in pure alcohol.
Insert the slider together with the
bulb socket.
Hold the new bulb in a cloth and push it into position.
Push the slider carefully but complete- ly back along the rail to the stop.
Filters • The
• You can insert other ∅ 50 mm filters if you need them:
Pull out the slider.
Insert the ∅ 50 mm filter.
Push the slider carefully but
completely back along the rail to the stop.
Starting up The
For
Set the voltage selector to 115.
Fit the enclosed two 315 mAt fuses in the fuse holder instead of the two 160 mAT fuses (see the section "Changing fuses" on the right).
Connect the power cable to the socket of the base and to the power source.
24 | Leica M Stereomicroscopes – Assembly |