Large swinging-arm and table-clamp stands
Safety ring ➜ column
Safety ring ➜ column
Limitation
Drive housing ➜ horizontal arm
The safety ring secures the instrument and must always be positioned beneath the horizontal arm.
Position the safety ring with the screw head uppermost and towards you, and slide it down the column.
Retighten the clamping screw.
Lower the horizontal arm onto the safety
ring, with its end pointing towards you.
Retighten the clamping screw.
The screw head must engage the notch. It limits the lateral movement. It may not be removed, and neither may the safety ring be milled down.
If the horizontal arm is swung out beyond the limit, the instrument may fall over.
This section describes the assembly of the focusing drive with
Release the clamping lever.
Tilt the carrier rod downwards.
Retighten the clamping lever.
Slide the drive housing with
Retighten the clamping lever.
The clamping levers can be positioned as required.
Pull out the clamping lever along its axis and turn it.
30 | Leica M Stereomicroscopes – Assembly |