4.Engage the highest magnification position.
5.Optimize the focus using the focusing drive.
C Adjusting | 6. Engage the lowest magnification |
dioptric settings | position. |
7. Do not look into the eyepieces.
Rotate the eyepieces counterclockwise in the "+" direction as far as the stop (+5 diopters).
8. Look into the eyepieces
Slowly rotate each eyepiece individually in the
D Inspecting 9. Engage the highest magnification parfocality position.
10.Refocus if necessary.
11.Slowly move the magnification changer from the lowest magnification to the highest.
•The image sharpness must remain constant (parfocal). Otherwise, repeat the procedure.
Leica M Stereomicroscopes – Operation | 21 |