| TROUBLESHOOTING & REPAIR |
INPUT RECTIFIER TEST (CONTINUED)
FIGURE F.4 – INPUT RECTIFIER TEST
INPUT | A |
RECTIFIER | B |
| C |
NEG
POS (+)
Return to Section TOC
Return to Section TOC
Return to Master TOC
Return to Master TOC
TEST PROCEDURE
1.Remove input power to the Power Wave 455M.
2.Using the 3/8” nut driver, remove the case top.
3.Perform the Capacitor Discharge Procedure.
4.Locate the Input Rectifier and lead locations. Refer to Figure F.4.
NOTE: Some silicone sealant may have to be removed from the input rectifier termi- nals. The G.E. silicone or equivalent should be replaced when test is com- plete.
5.With the phillips head screw driver remove the positive and negative leads from the rec- tifier.