Diagnostic Error Messages
The controller runs background tests during normal operation. If a problem with the background tests occurs, an error message is shown (flashing) in the Display. These are called Diagnostic Error Messages (Table 3).
Table 3 - Diagnostic Error Messages
|
|
| Display | Corrective Action | |||
|
|
|
|
|
|
|
|
|
|
| rPLC | 1. Turn Blower Switch OFF for 2 | |||
|
|
| Flashes on the display | minutes, then Turn Blower Switch | |||
|
|
| ON. | ||||
|
|
| alternately with | ||||
|
|
|
|
| |||
|
|
| temperature indicating | 2. Replace the controller if the | |||
|
|
| that the controller’s | ||||
|
|
| problem does not clear after repeating | ||||
|
|
| internal diagnostics test | ||||
|
|
| action 1 above three times. | ||||
|
|
| has failed. | ||||
|
|
|
|
| |||
|
|
|
|
|
|
|
|
|
|
| rSET | 1. Turn Blower Switch OFF for 2 | |||
|
|
|
|
|
| ||
|
|
| Oven temperature has |
|
| ||
|
|
|
|
| minutes, then Turn Blower Switch | ||
|
|
| not reached at least |
|
| ON. | |
|
|
| 200˚F(93˚C) within 15 |
|
| 2. If problem does not clear, call | |
|
|
| minutes of operation; |
|
| ||
|
|
| usually a result of the |
|
| the Middleby Cooking Systems | |
|
|
| burner not lighting. |
|
| Group (MCSG) for assistance. | |
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| 1. Check that the thermocouple is | |
|
|
|
|
|
| connected to terminals 7 & 8 of the | |
|
|
|
|
|
| temperature controller. | |
|
|
| OPEn | 2. If connections are good, then check | |||
|
|
| the continuity of the thermocouple at | ||||
|
|
|
|
|
| ||
|
|
| Caused by an | room temperature. The resistance | |||
|
|
| open circuit in the | reading should be in the 9 - 20 Ohm | |||
|
|
| thermocouple circuit. | range. Remove & replace an open | |||
|
|
|
|
|
| thermocouple. | |
|
|
|
|
|
| 3. If the thermocouple checks good, | |
|
|
|
|
|
| call MCSG for assistance. | |
|
|
|
|
|
|
| |
|
|
| r TC | 1. Check the polarity of the | |||
|
|
| thermocouple leads; terminal 7 | ||||
|
| Reverse polarity of the | RED wire & terminal 8 (+) WHITE | ||||
|
| thermocouple leads. | wire. | ||||
|
|
|
|
|
| 2. If polarity connections are | |
|
|
|
|
|
| correct replace thermocouple. | |
|
|
|
|
|
|
|
|