Applications and loading
The spray arms must not be blocked
by items which are too tall for the
basket or which hang down. Rotate
the arms manually to test.
Important
Make sure that the spring adapter for
water connection engages correctly
when a basket, injector or mobile unit is
inserted into the machine. It must be 4 -
5 mm higher than the water connection
inlet in the roof of the cabinet.
If it is not, adjust the adapter accord-
ingly.
Loosen the lock ring.
Push up the adapter (4-5 mm higher
than the water connection inlet) and
tighten the lock ring.
Select baskets and inserts which are
appropriate for the application.
Loading example:
E 380 "Intermiel"
Injector mobile unit (with TA connec-
tion)
For narrow necked glassware, com-
plete with 15 nozzles 4x1 mm/160 mm
high, 18 nozzles 6x1 mm/220 mm high
with 33 "Mielava" clips.
O 175 Injector mobile unit
(with (TA) Drying unit connection)
(top basket) for direct injection of nar-
row necked glassware, 34 nozzles 4x1
mm/ 160 mm high with "Mielava" clips.
Please note:
powder detergent can only be placed
in the inner side of the door when using
the O 175 unit .
Applications and loading
14