Plumbing

AD connection to purified water, pressurised, (1.5 - 10 bar)

The machine is delivered from the fac- tory ready for connection to a press- urised system, 1.5 - 10 bar.

The AD pressure tested hose, marked H2O pure, with 3/4" thread must be connected to the on-site AD stopcock for purified water.

Connection to purified water non- pressurised (0 - 1.5 bar)

For connection to a supply of 0 - 0.3 bar, or 0.3 - 1.5 bar the machine must be converted, if not supplied converted ex-works. An “ADP” booster pump can only be fitted. This is only possible when the machine has no DHK steam condensor. As this work requires specialised knowledge, it should only be carried out by a Miele approved engineer.

The outlet of the purified water con- tainer (non-pressurised) must be at least at the height of the upper edge of the machine. (see "Installation instruc- tions").

Connect the non-pressurised AD water hose to the purified water con- tainer

If the AD connection to purified water is not being used (or if it should not be used) then the ma- chine must be converted by a Miele trained or approved engineer.