NI PXI-5422 Calibration Procedure
Figure 3.
NI PXI-5422 Front Panel Connectors
PFI 0
PFI 1
NI PXI-
5422
16-Bit 200 MS/s AWG
ACCESS ACTIVE
CLK
IN
CH 0
DIGITAL DAT
A & CONTROL
Contents
Main
CALIBRATION PROCEDURE
NI PXI-5422
Contents
Conventions
Introduction
What Is Calibration?
Self-Calibration
External Calibration
Why Should You Calibrate?
How Often Should You Calibrate?
Software and Documentation Requirements
Software
Page
Documentation
Self-Calibration Procedures
MAX
FGEN Soft Front Panel
NI-FGEN
External Calibration Options
Complete Calibration
Figure 1. Complete Calibration Programming Flow
Optional Calibration
Figure 2. Optional Calibration Programming Flow
External Calibration Requirements
Test Equipment
Test Conditions
External Calibration Procedures
Writing Your Calibration Procedure
Calibration Procedures in LabVIEW
Calibration Procedures in LabWindows/CVI
Calibration Procedures in C
Verifying NI PXI-5422 Specifications
Figure 3. NI PXI-5422 Front Panel Connectors
National Instruments Corporation 15 NI PXI-5422 Calibration Procedure
NI PXI-5422
DIGITAL DATA & CONTROL
Verifying the Oscillator Frequency Accuracy
Page
Verifying the DC Gain and Offset Accuracy
Verifying the Main Analog Path Gain
Page
Page
Page
Table 4. Values for Verifying the Gain of the Main Analog Path
Table 4. Values for Verifying the Gain of the Main Analog Path (Continued)
Page
Verifying the Main Analog Path Offset
Table 5. Values for Verifying the Offset of the Main Analog Path
Table 5. Values for Verifying the Offset of the Main Analog Path (Continued)
Page
Verifying the Gain of the Direct Path
Table 6. Values for Verifying the Gain of the Direct Analog Path
Page
Adjusting the NI PXI-5422
Page
Figure 4. NI PXI-5422 External Calibration Procedure
Initializing the External Calibration Session
Adjusting the Analog Output
Initializing Analog Output Calibration
Page
Adjusting the Main Path Pre-Amplifier Offset
Page
Page
Page
Adjusting the Main Path Pre-Amplifier Gain
Page
Page
Page
Adjusting the Main Path Post-Amplifier Gain and Offset
Page
Page
Adjusting the Direct Path Gain
Page
Page
Page
Adjusting the Oscillator Frequency
Page
Page
Adjusting the Calibration ADC
Page
Page
Page
Page
Closing the External Adjustment Session
Calibration Utilities
MAX
FGEN SFP
NI-FGEN
Where to Go for Support