NI PXI-5422 Calibration Procedure 16 ni.com
Verifying the Oscillator Frequency AccuracyThis test verifies the frequency accuracy of the oscillator on the
NI PXI-5422. The verification involves generating a 10 MHz sine wave
with the NI PXI-5422 and measuring the sine wave frequency with one of
the instruments from Table 2.
To verify the frequency accuracy of the oscillator on the NI PXI-5422,
complete the following steps:
1. Connect the NI PXI-5422 CH 0 front panel connector to the instrument
measuring the frequency accuracy with a male BNC to female SMB
cable.
2. Call niFgen_init (niFgen Initialize VI) using the following
parameters:
•resourceName: The name of the device that you want to verify.
You can find this name under Devices and Interfaces in MAX.
•IDQuery: VI_TRUE
•resetDevice: VI_TRUE
•vi: A pointer to a ViSession. The variable passed by reference
through this parameter receives the value that identifies the
session created by this function. This value acts as the session
handle and is passed as the first parameter to all subsequent
NI-FGEN functions.
3. Call niFgen_SetAttributeViReal64 to set the sample rate
(niFgen property node: Arbitrary Waveform Output»Sample Rate)
using the following parameters:
•vi: The session handle returned from niFgen_init
•channelName: "0"
•attributeID: NIFGEN_ATTR_ARB_SAMPLE_RATE
•value: 200000000
4. Call niFgen_SetAttributeViReal64 to set the gain (niFgen
property node: Arbitrary Waveform Output»Arbitrary Waveform
Gain) using the following parameters:
•vi: The session handle returned from niFgen_init
•channelName: "0"
•attributeID: NIFGEN_ATTR_ARB_GAIN
•value: 1
Note You can adjust this value based on which measuring device you use.