© National Instruments Corporation 55 NI PXI-5422 Calibration Procedure
c. Call niFgen_InitiateGeneration (niFgen Initiate
Generation VI) to initiate the waveform generation using
the following parameter:
vi: The session handle returned from niFgen_InitExtCal
d. Measure the frequency of the generated waveform. This value is
the measured frequency.
14. Call niFgen_AbortGeneration (niFgen Abort Generation VI) to
abort the waveform generation using the following parameter:
vi: The session handle returned from niFgen_InitExtCal
Adjusting the Calibration ADC
The NI PXI-5422 has an onboard calibration ADC that is used during
self-calibration. Adjusting the calibration ADC involves characterizing the
gain and offset associated with this ADC so that performing self-calibration
results in an accurately calibrated device.
1. Call niFgen_InitializeCalADCCalibration (niFgen Initialize
Cal ADC Calibration VI) using the following parameter:
vi: The session handle returned from niFgen_InitExtCal
2. Call niFgen_WriteBinary16AnalogStaticValue (niFgen Write
Binary 16 Analog Static Value VI) to set the main DAC value using
the following parameters:
vi: The session handle returned from niFgen_InitExtCal
channelName: "0"
value: 0
3. Call niFgen_SetAttributeViInt32 to set the analog path value
(niFgen property node: Output Attributes»Analog Path) using the
following parameters:
vi: The session handle returned from niFgen_InitExtCal
channelName: "0"
attributeID: NIFGEN_ATTR_ANALOG_PATH
value: NIFGEN_VAL_FIXED_LOW_GAIN_ANALOG_PATH
4. Call niFgen_SetAttributeViInt32 to set the gain DAC value
(niFgen property node: Calibration»Gain DAC Value) using the
following parameters:
vi: The session handle returned from niFgen_InitExtCal
channelName: "0"
attributeID: NIFGEN_ATTR_GAIN_DAC_VALUE
value: 1700