© National Instruments Corporation 37 NI PXI-5422 Calibration Procedure
8. Call niFgen_SetAttributeViReal64 to set the post-amplifier
attenuation (niFgen property node: Calibration»Post-Amplifier
Attenuation) using the following parameters:
•vi: The session handle returned from niFgen_InitExtCal
•channelName: "0"
•attributeID: NIFGEN_ATTR_POST_AMPLIFIER_ATTENUATION
•value: 0
9. Call niFgen_SetAttributeViReal64 to set the output impedance
(niFgen property node: Basic Operation»Output Impedance) using
the following parameters:
•vi: The session handle returned from niFgen_InitExtCal
•channelName: "0"
•attributeID: NIFGEN_ATTR_OUTPUT_IMPEDANCE
•value: 50
10. Call niFgen_SetAttributeViBoolean to enable the analog output
(niFgen property node: Basic Operation»Output Enabled) using
the following parameters:
•vi: The session handle returned from niFgen_InitExtCal
•channelName: "0"
•attributeID: NIFGEN_ATTR_OUTPUT_ENABLED
•value: VI_TRUE
11. Call niFgen_Commit (niFgen Commit VI) to commit the attribute
values to the device using the following parameter:
•vi: The session handle returned from niFgen_InitExtCal
Adjusting the Main Path Pre-Amplifier Offset1. Call niFgen_SetAttributeViInt32 to set the analog path value
(niFgen property node: Output Attributes»Analog Path) using
the following parameters:
•vi: The session handle returned from niFgen_InitExtCal
•channelName: "0"
•attributeID: NIFGEN_ATTR_ANALOG_PATH
•value: NIFGEN_VAL_FIXED_LOW_GAIN_ANALOG_PATH