© National Instruments Corporation 9 NI PXI-5422 Calibration Procedure
Figure 1. Complete Calibration Programming Flow
Adjust
(Cal Dates and
Temperatures
Updated)
Verify
Document
Post-Adjustment
Results
Meets
Calibration
Test Limits?
Review
Verification/
Adjustment
Procedure or
Return Device
Verify
Self-Calibrate
Calibration/
Verification
Complete
Yes No
Document
Pre-Adjustment
Results