Omega CNI8DV, CNI16D, CNI8C, CNI8DH, CNI32 manual 2 Conversion Table, Range, Conversion Number

Models: CNI16 CNI32 CNI8DH CNI8C CNI16D CNI8 CNI8DV

1 72
Download 72 pages 49.28 Kb
Page 28
Image 28
Table 3.2 Conversion Table

Conversion number is a coefficient of conversion between input values and real full display range (10000 counts shown as 9999). See Table 3.2 below for proper conversion number.

Table 3.2 Conversion Table

RANGE

 

 

CONVERSION NUMBER

100 mV

 

 

10000 / (100 x 1) = 100

1 V

 

 

10000 / (1000 x 1) = 10

10 V

 

 

10000 / (1000 x 10) = 1

0 -20 mA

 

 

10000 / (20 x 1) = 500

Example

=

 

 

0 - 1 V

=

0

- 100.0

In 1

=

0

 

Rd 1

= 0

 

Inp 2

=

9999

Rd 2

= 100.0

24

Page 28
Image 28